DPI 104Specifications

Pressure Bar Graph
	In addition to the numeric pressure indication, the
	DPI 104 LCD display contains a round twenty segment
	bar graph to provide the user with a quick visual
	estimation of pressure from 0 to 100% full scale output.
	The bar graph increments represent 5% of the userselectable
	range.
	Display Update
	Two times per second
	Minimum/Maximum Values
	The minimum and maximum pressure values can
	be displayed on the DPI 104. This function can be
	enabled/disabled, and reset by the user.
	Switch Test
	The DPI 104 features a switch test function that will
	capture and display the open and closed values from an
	external pressure switch. Maximum switch impedance
	200 .
	Voltage Output
	The DPI 104 can be programmed to provide a 0 to 5 V
	output signal that can be configured as directly proportional
	to the pressure shown in the display or set to
	a fixed value. The voltage output mode provides 0.1%
	FS accuracy from 50 mV to 5 V.
	Alarm Output
	The alarm output consists of an open drain field effect
	transistor (maximum current 250 mA, maximum voltage
	24 VDC)
	Adjustable Mounting Positions
	For added convenience, the DPI 104 housing will rotate
	320° around the pressure fitting and the faceplate can
	be rotated in any orientation for optimum visibility.
	Menu Lock
	To guard against unauthorized menu access, the DPI 104
	features a menu and tare lock function
	RS232 Interface
	Serial communications are provided to allow transfer of
	data to a PC with the optional serial lead (IA090-1-VO).
	Using this link all menu commands and display data are
	available via an ASCII command set or the optional SiCal
	Pro Software.
	Network Capability
	Up to 99 DPI 104 instruments can be connected together
	in a daisy chain-configured network.
	DPI 104
	Specifications
	Universal Pressure Module (UPM) Capability
	UPM modules can be connected to the DPI 104 to
	enhance the accuracy and extend the pressure range.
	Battery
	The DPI 104 is supplied with a 9V alkaline battery, type
	MN1604. For increased performance a 9V lithium battery
	(not supplied by GE) is suggested.
	Pressure Port
	• 1/4 NPT or BSP male for units to 10,000 psi (700 bar)
	• 9/16 x 18 UNF male cone connection 15,000 psi
	(1000 bar) and 20,000 psi (1400 bar) units
	Enclosure
	• Case material: ABS/PC plastic sealed to Type 4/IP65
	• Wetted parts: All stainless steel (316) or Inconel welded
	pressure cavity for compatibility with aggressive
	media.
	European Compliance
	CE marked
	SiCal Pro Software
	This software package allows the user to control the
	DPI 104 remotely through a virtual instrument panel on
	the computer screen. The calibration data can be logged,
	viewed and printed in graphical format or as a
	calibration certificate.
	System Requirements
	Minimum Intel Pentium with Windows® 95 or higher.